Location:Home > Products > Furnace front molten iron analyzer > KD-TS8 furnace molten iron quality management instrument
Description:The KD-TS8-type pre-furnace iron quality management instrument adopts the original element, with high measurement accuracy and stable use. Calculate the carbon proportion (CE%), carbon content (C%), silicon content (SI%), ballization rate (SG%), and anti -tensile strength (RM) based on the characteristic points of temperature changes.
Product Category
Furnace front molten iron analyzer
Infrared carbon and sulfur analyzer
Furnace front carbon silicon analyzer
High frequency infrared carbon sulfur analyzer
Direct reading spectroscopic analyzer
Molten iron quality management instrument
Hardness tester, tensile tester
Quality management instrument for molten iron in front of the furnace
Related articles
Related ArticlesIntroduction
The KD-TS8-type pre-furnace iron quality management instrument adopts the original element, with high measurement accuracy and stable use. The pre -furnace heat analysis is based on the solidification temperature of the formation of the cast iron tissue into the object of the test. Calculate the carbon proportion (CE%), carbon content (C%), silicon content (SI%), ballization rate (SG%), and anti -tensile strength (RM) based on the characteristic points of temperature changes. The detection time of the iron water quality management instrument in front of the furnace is about 1 minute and 30 seconds. The detection is completed within the time when the iron water can be waited, and the iron water composition is adjusted according to the automatic calculated carbon increase agent, silicon iron, and scrap steel. Iron water is poured after the ingredients.
KD-TS8-type iron water quality management instrument main performance:
1. Instrument specifications:
①, original temperature transmitter, industrial control motherboard, 21 -inch LED real color widescreen display
②, power supply: AC 220V 50Hz power: 100W chassis size: 480 × 260 × 400mm
2. Measurement function:
①The iron and water brand, carbon volume, carbon content, silicon content, over -cooling, ballization rate, tensile strength, etc.
②, the setting of the target material, the setting of iron water weight
③ The amount of automatic calculation of carbon increase agents, silicon iron, and scrap steel
④ Automatic drawing carbon, silicon, carbon proper amount, early crystal temperature, common crystal temperature, and overdue change detection line
⑤ Preservation and reproduction of data data
3. Setting function: You can set 10 inspection lines, 10 target components, inspection lines and target materials linkage
4. Make -up function: Calculate and increase the amount of target components, iron water weight, and income from the measurement results
5. Record function: According to the measurement time order, you can record 10,000 measured results, and you can reproduce the process
6. Browse function: The trend distribution map of the main measurement parameters can be browsed so that data analysis can be analyzed
7. Output function: The measurement record can be transferred to other computers to check through the U disk
The main parameters of the iron water quality management instrument of KD-TS8 furnace:
Measurement name measurement range measurement precision
TEMP 0-1370 ℃ F.S ± 0.01%
CE % 2.20 % -5.00 % ± 0.05 %
C % 2.00 % -4.80 % ± 0.05 %
Si % 0.20 % -5.50 % ± 0.10 %
SG % 62 % -95 % ± 10.0 %
RM HTL00 -HT350
Product consultation
Welcome to add our WeChat account to learn more information